Maintenance Tests
9650-0807-01 Rev. A
Propaq MD Service Manual
31
SpO
2
Test
Equipment
Fluke Index 2 SpO
2
Simulator (or equivalent), Masimo
®
Rainbow
®
SpO
2
finger probe
Test Setup
Disconnect any ECG cable or IBP probes from the Propaq MD unit.
Do This...
Observe This...
Pass/Fail
1.
Power on the Propaq MD unit without the
SpO
2
sensor attached.
Verify that the
SpO
2
numeric
displays INITIALIZING, then
goes blank.
2.
Connect the
SpO
2
sensor to the Propaq MD
unit.
Verify that
SpO
2
numeric
displays SEARCHING.
3.
Set the simulator to 70%
SpO
2
and 60 PPM.
Attach the
SpO
2
sensor to the simulator.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
67% and 73%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
o
o
o
o
4.
Set the simulator to 80%
SpO
2
and 60 PPM.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
77% and 83%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
o
o
o
o
5.
Set the simulator to 90%
SpO
2
and 60 PPM.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
87% and 93%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
o
o
o
o
6.
Disconnect the
SpO
2
simulator.