beautypg.com

Test – ZOLL Propaq MD Rev A User Manual

Page 35

background image

Maintenance Tests

9650-0807-01 Rev. A

Propaq MD Service Manual

31

SpO

2

Test

Equipment

Fluke Index 2 SpO

2

Simulator (or equivalent), Masimo

®

Rainbow

®

SpO

2

finger probe

Test Setup

Disconnect any ECG cable or IBP probes from the Propaq MD unit.

Do This...

Observe This...

Pass/Fail

1.

Power on the Propaq MD unit without the

SpO

2

sensor attached.

Verify that the

SpO

2

numeric

displays INITIALIZING, then
goes blank.

2.

Connect the

SpO

2

sensor to the Propaq MD

unit.

Verify that

SpO

2

numeric

displays SEARCHING.

3.

Set the simulator to 70%

SpO

2

and 60 PPM.

Attach the

SpO

2

sensor to the simulator.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

67% and 73%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

4.

Set the simulator to 80%

SpO

2

and 60 PPM.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

77% and 83%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

5.

Set the simulator to 90%

SpO

2

and 60 PPM.

Verify that after a brief period of
initialization, the

SpO

2

numeric

displays an

SpO

2

value between

87% and 93%.

Verify that the displayed pulse
rate reads 60 +/- 3 bpm.

o

o

o

o

6.

Disconnect the

SpO

2

simulator.