beautypg.com

Advisory test – ZOLL Propaq MD Rev A User Manual

Page 30

background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

26

9650-0807-01 Rev. A

Advisory Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent), ECG simulator

Do This...

Observe This...

Pass/Fail

1.

Connect the multifunction cable (MFC) to
the Analyzer.

2.

Turn on the Propaq MD unit and connect the
pads.

The unit displays Pads on the top
left of the screen.

o

o

3.

On the Analyzer, select VF (ventricular
fibrillation), then press the Analyze quick
access key.

The Propaq MD unit displays:

ANALYZING

SHOCK ADVISED

o

o

4.

On the Propaq MD unit, press the
CHARGE button to activate, then press the
CHARGE button again.

The unit charges to the selected
energy level.

o

o

5.

On the Propaq MD unit, press the Shock
button.

Verify that the shock count is
displayed at the top of the screen.

o

o

6.

On the Analyzer, select NSR (normal sinus
rhythm), then press the Analyze quick
access key.

The Propaq MD unit displays:

ANALYZING

NO SHOCK ADVISED

o

o