5.

On the Propaq MD unit, do the
following:

Press the More (

) then the

Setup (

) quick access keys.

Use the navigation keys to
select Supervisor>Display/
Configuration>Device Info
and select Restore All Factory
Defaults
.

Note: You need a supervisor
passcode to enter the Supervisor
menu.

6.

Press the I Measure switch on the
Battery Current Test Fixture.

Observe the current measurement on
the DMM, and verify that the reading
is between 0.47mA-1.25mA.