Overview

This appendix includes simulators and settings that may be used to assess the performance of the NIBP module.
Readings should be obtained within +/- 10 mmHg of the expected measurement result. The following simula-
tors are used in this chapter:

Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump

Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump2

Fluke Biomedical BP Pump/Dynatech Nevada Inc. CuffLink

Clinical Dynamics Smart Arm

NIBP Simulator Performance

FastBP

A dual lumen hose is required for FastBP.When you use FastBP, the reading is obtained during cuff inflation.

StepBP

StepBP is compatible with both single and dual lumen hoses.When you use StepBP, the reading is obtained
during step deflation.