Treatment report test – ZOLL Propaq MD Rev A User Manual
Page 34
C
HAPTER
1
M
AINTENANCE
T
ESTS
30
9650-0807-01 Rev. A
Treatment Report Test
Equipment
Fluke Impulse 7000 DP Analyzer (or equivalent)
Do this...
Observe this...
Pass/Fail
1.
Connect the multifunction cable (MFC) to the
Analyzer, and turn on the Analyzer.
2.
Turn on the Propaq MD unit, and press the
following quick access keys:
a)
b)
c)
The Clear Log window
displays.
3.
Select Yes.
The unit displays a clearing
the log message.
o
o
4.
Press the CHARGE button, select 200 joules,
then press the CHARGE button again.
The unit starts to charge, a
charging tone sounds, and the
Shock button illuminates.
o
o
5.
Press the Shock button.
The unit discharges the shock.
o
o
6.
Wait 20 seconds, press the Treatment quick
access key(
), and select the first code
marker.
7.
Wait 15 seconds and turn off the unit, then wait
2 minutes and turn on the unit.
8.
On the Propaq MD unit, press the following
quick access keys:
a)
b)
c)
The Treatment Summary
window displays.
o
o
9.
Highlight the second event and press Select.
Verify that a green check mark
displays next to the file.
o
o
10.
Select Print Treatment Summary.
Verify that the recorder prints
out the defib event and the
code marker snapshot.
o
o
11.
Turn off the Propaq MD unit.
LOG
R
LOG