beautypg.com

Treatment report test – ZOLL Propaq MD Rev A User Manual

Page 34

background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

30

9650-0807-01 Rev. A

Treatment Report Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent)

Do this...

Observe this...

Pass/Fail

1.

Connect the multifunction cable (MFC) to the
Analyzer, and turn on the Analyzer.

2.

Turn on the Propaq MD unit, and press the
following quick access keys:

a)

b)

c)

The Clear Log window
displays.

3.

Select Yes.

The unit displays a clearing
the log message.

o

o

4.

Press the CHARGE button, select 200 joules,
then press the CHARGE button again.

The unit starts to charge, a
charging tone sounds, and the
Shock button illuminates.

o

o

5.

Press the Shock button.

The unit discharges the shock.

o

o

6.

Wait 20 seconds, press the Treatment quick

access key(

), and select the first code

marker.

7.

Wait 15 seconds and turn off the unit, then wait
2 minutes and turn on the unit.

8.

On the Propaq MD unit, press the following
quick access keys:

a)

b)

c)

The Treatment Summary
window displays.

o

o

9.

Highlight the second event and press Select.

Verify that a green check mark
displays next to the file.

o

o

10.

Select Print Treatment Summary.

Verify that the recorder prints
out the defib event and the
code marker snapshot.

o

o

11.

Turn off the Propaq MD unit.

LOG

R

LOG