Heart rate alarm test – ZOLL Propaq MD Rev A User Manual
Page 21
Maintenance Tests
9650-0807-01 Rev. A
Propaq MD Service Manual
17
Heart Rate Alarm Test
Equipment
Fluke Impulse 7000DP Analyzer
Do this...
Observe this...
Pass/Fail
1.
Press the Leads quick access key
(
)
on the Propaq MD unit and
select lead II.
II displays next to the Leads quick
access key.
o
o
2.
On the Propaq MD unit:
•
Press the More quick
access key (
) key to
access the second set of
quick access keys.
•
Press the Alarm quick
access key (
).
•
Press the Limits softkey.
The Alarm Settings menu displays.
o
o
3.
On the Propaq MD unit, select HR/
PR Alarm in the Alarm Settings
menu and use the navigation keys
to do the following:
•
Set the Upper Limit Enable
to ON.
•
Set the Lower Limit
Enable to ON.
•
Set the Lower limit to 35.
•
Set the Upper limit to 150.
Use the navigation keys to exit the
Alarm Settings menu.
4.
On the Propaq MD unit, press the
Alarm Cancel quick access key
(
).
The red alarm suspension timer
displays at the top of the screen next
to the message area.
5.
Remove a lead wire from the
ECG
simulator
.
T
he ECG LEAD FAULT message
displays.
o
o
6.
On the Propaq MD unit, press
.
The Alarm Suspend key flashes
.
7.
Reattach ECG Lead wire to
ECG
simulator
.
The ECG LEAD FAULT message no
longer displays.
o
o
8.
Set simulator to 30 BPM or lower.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM message displays.
9.
On the Propaq MD unit, press
.
The Alarm Suspend key flashes.
The
red alarm suspension timer displays
at the top of the screen next to the
message area.
o
o
10.
Set the simulator to 120 BPM.
11.
On the Propaq MD unit, press
.
Alarm is enabled. The alarm suspend
key flashes. The alarm suspension
timer disappears.
o
o
I, II,
III...