beautypg.com

Leads test – ZOLL Propaq MD Rev A User Manual

Page 19

background image

Maintenance Tests

9650-0807-01 Rev. A

Propaq MD Service Manual

15

Leads Test

Equipment

3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer

Note:

Test each cable separately.

Test Setup

Connect the lead wires appropriate for each test to the

Fluke Impulse 7000DP Analyzer

or

equivalent, then connect the ECG cable to the Propaq MD unit.

Do this...

Observe this...

Pass/Fail

1.

Turn on the Propaq MD unit.
Note: Make sure the lead is not
displayed.

No ECG LEAD FAULT message
displayed.

o

o

2.

Disconnect one lead from the
simulator.

The ECG LEAD FAULT message displays
within 3 seconds (if configured).

o

o

3.

Reconnect the lead. Repeat step 2
with the remaining leads.

Wait for ECG LEAD FAULT message to
clear from the display (if configured).

o

o

4.

Repeat steps 2 and 3 for 5 lead and 12
lead cables.

Note: If heart rate alarm sounds, on the
Propaq MD unit:

Press the More quick access key (

) key

to access the second set of quick access
keys.

Press the Alarm quick access key (

).

Press the Alarm Cancel quick access key

(

).

o

o