Maintenance Tests

9650-0807-01 Rev. A

Propaq MD Service Manual

35

IBP Test

Tools Needed

IBP Simulator (with accuracy better than +/- 1 mmHg), or equivalent

Setup

Connect the test cables to IBP patient simulator.

Do This...

Observe This...

Pass/Fail

1.

Connect the IBP patient simulator to the
Propaq MD P1 channel.

2.

Set the IBP patient simulator to 0 mmHg.

3.

On the Propaq MD unit, press > IBP.

4.

On the Propaq MD unit, Press > P1->0<-. Display of 0/0(+/-2) in the P1 box

at the bottom of the display (P1 is
zeroed).

The waveform for channel 1 is
present.

o

o

o

o

5.

Exit the menu to see the waveform.

6.

Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).

Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).

o

o

7.

Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.

8.

Disconnect IBP cable.