Propaq MD Error Messages
9650-0807-01 Rev. A
Propaq MD Service Manual
55
NIBP TEMP OUT OF
RANGE
Module temperature is out of
its specified operating range.
Move the device to a cooler
or warmer environment.
NIBP WEAK PULSE
Too few pulses are detected
during the reading attempt.
Tighten the cuff and ensure
appropriate placement.
PACER CURRENT FAULT
Lack of pacer current.
Replace the defibrillator
board module.
PACER DEVICE FAILURE
Generation of pacer pulse
failed.
Replace the defibrillator
board module.
PRINTER OFFLINE
Printer module is off-line.
If it remains off line, try
checking the connectors or
power to the printer. If
problem persists, replace
the printer module.
PRINTER OUT OF PAPER
If there is paper in the
device, probably a paper
sensor problem.
Replace the printer module.
PRINTER PLATEN OPEN
If printer door is closed,
probably a problem with the
micro switch.
Replace the printer module.
PS DEFIB CHARGE
SHORT
One of the paddle set keys
are stuck.
Replace the paddle set or
MFC.
PS DEFIB ENERGY DOWN
Paddle set keys are stuck.
Replace the paddle set or
MFC.
PS DEFIB ENERGY DOWN
SHORT
Paddle set energy down
select button is stuck.
Replace the paddles.
PS DEFIB ENERGY UP
Paddle set keys are stuck.
Replace the paddle set or
MFC.
PS DEFIB FIRE SHORT
Paddle set keys are stuck.
Replace the paddle set or
MFC.
PS DEFIB RECORDER
SHORT
One of the paddle set keys
are stuck.
Replace the paddle set or
MFC.
QUICK ACCESS SW 1(to
SW 7) SHORT
One of the quick access
keys (key 1 to key 7) is
stuck.
Replace the front panel.
SCREEN CYCLE SHORT
Display/home key is stuck.
Replace the front panel.
Error Message
Explanation
Technical Action