Pacer Test
Equipment
Fluke Impulse 7000 DP Analyzer (or equivalent)
Do This...
Observe This...
Pass/Fail
1.
On the Analyzer, select the PACER
function per test equipment instructions.
2.
On the Analyzer, press F2 to select PACER
JACKS, then press F5 (Done) to measure
pulse.
3.
Connect the MFC connector to the
Analyzer.
4.
On the Propaq MD unit, apply power to the
unit using a battery or an AC adapter.
5.
On the Propaq MD unit, press the PACER
button.
6.
On the Propaq MD unit, use the navigation
keys to do the following:
•Set Rate to 30.
•Set Mode to Fixed.
•Set Output to 10.
•Select Start Pacer.
o
o
7.
On the Impulse 7000 DP Analyzer, record
the following:
•pulse rate
•pacer amplitude output
•pacer pulse width
•Pacer Rate PPM
(28.5ppm-31.5ppm)
•Pacer amplitude output
(5mA-15mA)
•Pacer pulse width (38ms to 42ms)
Note: If an ECG FAULT message
displays when leads are not
attached, it is normal.
8.
On the Propaq MD unit, use the navigation
keys to do the following:
•Set Rate to 180.
•Set Output to 140.
o
o
9.
On the Impulse 7000 DP Analyzer, record
the following:
•pulse rate
•pacer amplitude output
•pacer pulse width
•Pacer Rate PPM (162ppm-198ppm)
•Pacer amplitude output
(133mA-147mA)
•Pacer pulse width (38ms to 42ms)
10. Select Turn Pacer Off.