beautypg.com

Rear manifold removal / replacement – Ransburg Evolver 303 Dual Purge Solvent Robotic Applicator A12374-XXX User Manual

Page 51

background image

46

Evolver 303 Dual Purge Robotic Atomizers - Maintenance

Low Voltage Cable Removal

(Refer to Figure 18)

1. Disconnect low voltage cable [3] from rear of power

supply. For A10406-XX power supply, disconnect

circular connector from receptacle on rear of power

supply. For LECU5004-XX power supply, disconnect

wires from connector J3 at rear of MicroPak.

2. Loosen retaining ring [13] (see Figure 16) and

remove applicator from robot wrist.

3. Using 3/16" Allen wrench, remove set screw [2]

(see Figure 18).

4. Remove the low voltage cable [3] by pulling through

the robot arm and mounting plate [1].

5. To reinstall, insert power supply end of cable [3]

through mounting plate [1], pull through robot arm

and connect to junction box or power supply. For

connections to power supply, plug connector into re-

ceptacle (A10406-XX power supply) or connect wires

as shown in Figure 4 (LECU5004-XX power supply).

6. Align the timing mark of the plug on the appli-cator

end of the cable [3] with the mark on the mounting

plate [1] and tighten set screw [2].

Rear Manifold Removal / Replacement

Remove the rear manifold from the applicator (see

"Spray Applicator Disassembly" in the "Maintenance"

section). Inspect the valve weep ports on the mani-

fold body for leaks or contamination. If any paint is

visible in these areas, the valves need to be replaced

as shown in Figures 20 and 21.

Figure 19: Weep Port Locations

To replace the valves, use the valve removal tool

(A11922-00). First, engage the four (4) pins on the tool

to the corresponding four (4) hole pattern in the top

of the valve. Using a 1/2" (13mm) socket, end-wrench,

or adjustable wrench, remove the valve by turning

counter-clockwise, as shown in Figure 20.

Figure 20: Valve Removal

COUNTER-CLOCKWISE TO REMOVE

VALVE

A11922-00
VALVE REMOVAL TOOL

WEEP PORTS

AA-08-01.1