GF Signet 4150 Turbidimeter User Manual
Page 21
![background image](/manuals/331175/21/background.png)
21
Signet 4150 Turbidimeter
Troubleshooting
Signet 4150 Fault Detection
The Signet 4150 performs continuous diagnostic monitoring. There are three levels of fault detection: warnings, errors and failures.
Warning
A warning is simply a screen indication of a problem. The measurement is not interrupted and no alarms are activated.
• When the desiccant becomes saturated, a screen warning of DESC will appear.
Error
An error indicates a failure or a problem that usually can be corrected. These errors include:
• 4 to 20 mA loop open (MA)
• Bad calibration (CAL)
• Desiccant alarm activated and replacement required (DESC)
• Ultrasonic transducer is not making contact or the À ow through cuvette has been removed (CLN)
• Lamp out (LAMP)
The source lamps in the Signet 4150 are designed for long life. The IR lamp is rated for 10 years and the white light version is rated
for 7 years.
Replacing the lamp involves signi¿ cant disassembly and handling delicate components.
We recommend returning the unit to the factory for this service. If ¿ eld replacement is required, contact the factory for assistance.
If any of these conditions occurs, both relays will be set to the alarm state and the 4 to 20 mA output be held at the Error Level (ERLV)
selected in the Options Menu (if 4 to 20 mA is selected).
System FAIL Message
A failure is a system fault. This is NOT a problem that the operator can correct, and the unit must be returned to the factory for service.
If a failure occurs, the instrument will not function properly and will display the word FAIL on the lower row. Both alarm relays will be
activated and the 4 to 20 mA output will be held at the Error Level (ERLV) selected in the Options Menu (if 4 to 20 mA is selected).
Symptom
Possible Cause
Solution
Lower display shows NoMA
4 to 20 mA open loop warning when
4 to 20 mA is enabled in Options Menu
Check Loop wiring or turn off in Options Menu
Lower display shows DESC
High condensation
Check for leaks or replace desiccant pouch
Lower display shows LAMP
Lamp failed
Contact factory for assistance
Lower display shows FAIL
Major system fault
Return to factory for service
Readings are higher than expected
Bubbles in solution
Ensure that the drain vent is open and is not obstructed.
Apply back pressure.
Install an optional Stilling/Bubble Chamber
(4150-0003, 159 001 587).
Condensate or leaky cuvette
Check À ow through cuvette for condensate or leaks.
Flow-through cuvette dirty
Clean cuvette.
Instrument out of calibration
Recalibrate
Readings are erratic
Bubbles in solution
Ensure that the drain vent is open and is not obstructed.
Apply back pressure.
Install an optional Stilling/Bubble Chamber
(4150-0003, 159 001 587).
Debris in À ow-through cuvette
Clean debris from cuvette
Readings are lower than expected
Instrument out of calibration
Recalibrate.
Display shows "Err CAL"
Calibration error or bad standard used
Press the up key to recalibrate (see calibration) or
replace the standard.
Upper display À ashes
Sample Over-Range
Check sample. Sample may be too high to read.
Troubleshooting