Signet 4150 Turbidimeter
Valid Addresses
0001–00XXX
Input Status
These single-bit values are readable from the master. The data
will be returned with the lowest addressed input status in the LSB
of the data. Unused bits in the data will be set to 0.
Code
Name
Broadcast
0x01
Read Coil Status
No
0X05
Force Single Coil
Yes
Valid Command(s)
MSB
LSB
Bit
15
Bit
14
Bit
13
Bit
12
Bit
11
Bit
10
Bit
9
Bit
8
Bit
7
Bit
6
Bit
5
Bit
4
Bit
3
Bit
2
Bit
1
Bit
0
Format
16-bit word format
Address
Function
Default
00001
Offset added
False
00002
Flow alarm selected
False
00003
Access code enabled
False
00004
Ultrasonic cleaning enabled (if available)
True
00005
Desiccant set as error (True) or warnings (False)
True
De¿ nitions
Code
Name
Broadcast
0x02
Read Input Status
No
Valid Command(s)
MSB
LSB
Bit
15
Bit
14
Bit
13
Bit
12
Bit
11
Bit
10
Bit
9
Bit
8
Bit
7
Bit
6
Bit
5
Bit
4
Bit
3
Bit
2
Bit
1
Bit
0
Format
16-bit word format
Valid Addresses
10001–10XXX
Address
Function
10001
Instrument error
10002
Alarm 1 active
10003
Alarm 2 active
10004
Calibration error
10005
Desiccant error
De¿ nitions
Coils
These single-bit values are readable and changeable from the
master. The data will be returned with the lowest addressed coil in
the LSB of the data. Unused bits in the data will be set to 0.
True is 1 and False is 0.
Modbus RS-485 Output and Commands
If the connection is to the master on a RS-232 serial port, an RS-232 to RS-485 converter is required. If the connection to the master is
USB, an RS-485 to USB converter is required.