GF Signet 4150 Turbidimeter User Manual
Signet 4150 turbidimeter, English
![background image](/manuals/331175/1/background.png)
Signet 4150 Turbidimeter
3-4150.090
Rev N 11/12 English
*3-4150.090*
Speci¿ cations
Measurement Range
0 to 100.0 NTU or 0 to 1000.0 NTU
NTU = FNU = FTU
Accuracy
< 40 NTU:
±2% of reading or ±0.02 NTU
whichever is greater
> 40 NTU: ±5% of reading
Resolution
0.0001 NTU (below 10 NTU)
Response Time
Adjustable
Display
Multi-Line LCD with backlighting
Alarm Relays (2)
120 to 240 VAC, 2A Form C Relay
Current Output
Active 4 to 20 mA, 600
Digital Output
RS-485
Wetted Materials
• Tubing:
Vinyl
• Measuring Cuvette:
Borosilicate Glass
• Glass Washer Seal:
Silicone
• Pressure Regulator:
Polypropylene
FPM
316 Stainless Steel
Polyoxymethylene
(POM)
• Inlet Tube:
316 Stainless Steel
Maximum Inlet Pressure
Recommended 345 kPa (50 psi)
based on tubing connection provided
Pressure Regulator
Rated up to 200 psi
Flow Rate
0.1 L/m to 1 L/m
(0.026 GPM to 0.26 GPM)
Operating Temperature
1 °C to 50 °C (34 °F to 122 °F)
Sample Temperature Range 1 °C to 50 °C (34 °F to 122 °F)
Power Supply
100 to 240 VAC, 47 to 63 Hz, 80 VA
(12 to 24 VDC option available.
Contact
factory)
Insulation Rating
Double Insulated
Pollution Degree 2
Over voltage Category II
Altitude
2000 meters (6562 ft) maximum
Relative Humidity
Maximum 95% RH non-condensing
Enclosure Rating
Power Supply Box: NEMA 4X / IP66
Environmental Conditions Not recommended for outdoor use
Shipping Weight
2.5 kg (5.5 lb)
Standards and Approvals
CE
White Light version
Compliant to U.S. EPA 180.1
Infrared version
Compliant to ISO 7027
ETL Listed
UL 61010-1 and CSA C22.2
No.
61010-1
China
RoHS
For detailed information go to www.gfsignet.com
Description
The Signet 4150 Turbidimeter provides accurate and reliable water quality monitoring for municipal and industrial applications.
The 4150 uses the Nephelometric method to calculate the turbidity of a sample as it À ows through a viewing chamber. Models are
available that use white light technology as required by the U.S. EPA 180.1 standard, or select the models that use infrared technology
as required by ISO 7027.
Periodic calibration is mandatory with most turbidity systems, and the 4150 makes it fast and easy with sealed, reusable primary
calibration standards.
Two dry contact relays serve as high or low alarms, with programmable setpoints and time-based delays to prevent false alarms.
Additional features include a bright backlight for the display and a convenient holder for the cuvette during calibration.
An ultrasonic automatic cuvette cleaning system is standard for 0 - 100 NTU/FNU systems and optional for 0 - 1000 NTU/FNU systems.
Signet Turbidimeter
ENTER
Power Supply &
Wiring Terminals
English
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.