The Veris MSP pH Manager uses two electrodes for optimal data quality. If there is a difference of
0.75 or greater between the final electrode readings, an audible alarm will beep, informing the
operator of the erroneous reading.
To pause the data collection process at any time (but keep the same file), press the 1 key. Once
the sampling process has completed its cycle, it will disengage and the status text message will
indicate disengaged (press 1 to start cycling again). If the system no longer senses a speed signal
from the GPS, it will also disengage and return to Neutral. NOTE: do not depend on the GPS
speed signal for disengagement.
Before inspecting or working around any
component of the system, press the 1 key and
verify status of system is ‘Disengaged’ before
exiting the vehicle. Sporadic GPS signals may
simulate movement and initiate the cycling
process, resulting in possible entanglement and injury.
If TAP or DI are selected as the wash water type, the water baseline process will be repeated
every 40 cycles following the next engage press. If the pH during the cycling sequence does not
get within 0.5 of the baseline pH, an audible alarm will sound and the pH labels will blink. This is to
allow operator that the electrodes are not cleaning properly.