beautypg.com

Veris Technologies MSP3 - Operating Instructions User Manual

Page 25

background image

Pub. #OM17-MSP3

4-5


After the light reference has been read, the following will appear:

Figure 16

The reference data has now been stored. Once the system is restarted it is ready to start
mapping.

*NOTE: To ensure system is operating correctly always run a system check before
mapping a field.

There should be a difference of 100 or greater from dark reference to light reference for
each wavelength.

pH System Check


Calibrating pH electrodes
Enter menu option 2) pHsetup

Figure 17

Enter menu option 1) Calibration.

Figure 18

You will be asked for the ID of the electrode connected to channel 1. You may want to add an ID
number to the electrodes, for your own tracking purposes. Use the 1 and 2 keys to change the
number and 3 to confirm:

Figure 19

Repeat for electrode 2’s ID and press 3 to confirm.