beautypg.com

Veris Technologies MSP3 - Operating Instructions User Manual

Page 34

background image

Pub. #OM17-MSP3

4-14

The software will prompt for the electrodes to be inserted into pH buffer 4 solution; Slide cup with
pH 4 buffer solution onto electrode holder. Press 1 to continue with calibration or 2 to exit. Tips:
Don’t overfill solution. Cup only needs enough solution to immerse electrode tip and face. Don’t
reuse solutions.

Figure 43

The software will read the electrodes for 10 seconds, displaying the output (as it counts seconds):


Figure 44

After 10 seconds, the software will display the final pH reading and offer the options to Accept pH 4
buffer readings; Redo pH 4 buffer readings; or Cancel pH electrode calibration
. If the readings are
satisfactory, press Accept; if the readings are suspect, press Redo to return to re-do pH 4.

Figure 45