Veris Technologies MSP3 - Operating Instructions User Manual
Page 29
Pub. #OM17-MSP3
4-9
Figure 30
Select the type of water you are using to clean the electrodes between samples. The available
types are TAP, RO (reverse osmosis), or DI (de-ionized). Press 1 or 2 to cycle through the water
types, press 3 to continue to the next screen.
Tip: If you don’t want a baseline wash performed
every 40 cycles, use RO setting rather than Tap or DI (regardless of actual water being used).
Figure 31
Turning on the extra wash option will add 1.5 seconds of cleaning per cycle. The extra wash is
performed by stopping the shoe briefly during the cycle. Use this feature if you have noticed the
electrodes are not cleaning during the cycle. Press 1 or 2 to choose ON or OFF, press 3 to
continue. Tip: water usage will double if extra wash is used.