Veris Technologies MSP3 - Operating Instructions User Manual

Page 50

background image

Pub. #OM17-MSP3

4-30

Figure 75

2. Adjust electrode holder: with sampling mechanism raised completely, adjust electrode

holder to provide ½” (1.2 cm) clearance between it and sampling trough.

Figures 76a and 76b

3. Wash adjustment: Wash brackets should be parallel to sampling trough, with jets directly

beside electrode holder, jets should be 1” (2.5 cm) below electrodes; when electrodes are
properly aligned.


Figures 77a and 77b

Turn crank to raise
or lower electrode
holder

electrode
holder 1”
(2.5cm)
above wash
jets

Adjust scraper bracket
until cutting shoe
clears scraper when
sampler assembly is
raised completely.

½” (1.2 cm) clearance
between electrode holder
and trough liner