beautypg.com

Change of contact pulleys and springs, Figure 20, Replacing the pump hose – M&C TechGroup CSS-VC2 Operator's manual User Manual

Page 47

background image

14.1a ME

Gas sampling and gas conditioning technology

47

Figure 20

Replacing the pump hose

Unplug gas conditioning unit from the mains;

Remove hose connections on the pump;

Squeeze belt using the handles and turn the S-latch clockwise until it stops;

Remove the belt and pull the old pump tubing on the slip fittings out of the guides;

Squeeze the pressure rollers and see if spring tension exists; if not, change pressure springs
(see Section Fehler! Verweisquelle konnte nicht gefunden werden.);

Insert new pump hosewith slip fittings into the guides  of the belt;

W A R N I N G !

Only the use of original spare pump hose ensures proper
function.
Never lubricate the pump hose.
Check all parts for dirt before assembling the hose pump and
clean if necessary.

Insert belt complete with the new hose into the dovetail of the pump body; Squeeze the belt
at the handles and simultaneously turn the S-latch  counter clockwise until it clicks;

Switch pump on.


18.3.2 CHANGE OF CONTACT PULLEYS AND SPRINGS

Switch off the mains;

Unscrew nuts of the pump head (wrench size 5,5)  and remove snap ring  from motor

shaft;