Other fault and alarm messages, Maintenance – M&C TechGroup CSS-VC2 Operator's manual User Manual

Page 41

background image

14.1a ME

Gas sampling and gas conditioning technology

41


This prevents liquid pumped along to the analyser and thus damaging ! The triggered alarm (red LED)
must be acknowledged by pressing any key. If the error is not corrected, the error screen comes back
after 60s.

17.4

OTHER FAULT AND ALARM MESSAGES

Error message

Alarm Warning

LED

Error

Pump-Warn.: 1, 2

x

x

red
+
yello
w

Pump 1 and/or 2 off due to an alarm
(temperature, liquid)

Maintenance needed!

-

x

yello
w

Maintenance interval expired

Comm. Error

x

-

red

Error on connecting cable RJ45 between
CSS-VC.. and communication module

Flushing pressure not
present!

-

x

yello
w

Pressure for probe backflushing too low or
non-existent

Thermal Error: 2, 3

x

-

red

Thermocouple set and defective or not
connected or PT100 connected



18

MAINTENANCE

Before performing maintenance work, the system and process-specific safety measures must be
observed!

W A R N I N G !

Hazardous voltage. Pull the mains plug out before carrying out
work on the gas conditioning unit!