Gas sampling and gas conditioning technology
33
16.2
CONFIGURE THE SWITCHING ON AND OFF OF THE SAMPLE PUMPS
By 2 x pressing of the arrow key
and then 1 x pressing of the PRG key,
the following view
appears:
Here it is visible whether the pump is working or not. Using the arrow keys,
the pump to be
configured can be selected. By pressing the PRG key again,
the following view will appear:
Using the arrow keys,
the parameters to be
changed can be selected. By pressing the PRG key,
the parameter can now be changed with the arrow keys
and then confirmed with the PRG key
or
discarded with the ESC key
.
Here you select the parameter to be taken into consideration for switching off the pump.
- = Not taken into consideration/= x is taken into consideration
T1
=
Pump off with cooler temperature alarm (default setting x)
T2 + T3
=
Pump off with temperature alarms from external temperature control module or
Temperature controller for heated line in portable version (Art. No.. 01G6190)
(Factory setting x)
Ext
=
external pump control (with closed switch on) (Default setting -)
LA
=
Pump 1 resp. 2 off with liquid alarm 1 resp. 2 (x factory setting) At activated reset
(see 16.7) R occurs behind
– or x.
On
=
Pump is always on (overrides all other settings) (default setting -)
Off
=
Pump is always off (overrides all other settings) (default setting -)
RSp
=
Pump off during backflush programme (default setting x)