TEMPERATURE ALARMS FROM COOLER AND TEMPERATURE CONTROL
MODULE
Error message
Alarm
Warning
LED
Error
Lowtemp.: 1
x
-
red
Low temperature cooler
Hightemp: 1
x
-
red
High temperature cooler
T> Max: 1
x
red
Temperature sensor range exceeded
=> sensor defective
T x red Temperature sensor measuring range Lowtemp.: 2, 3 - x yellow Low temperature channel 2, 3 Hightemp.: 2, 3 x - red High temperature channel 2, 3 T> Max.: 2, 3 x red Temperature sensor range exceeded T x red Temperature sensor measuring range 17.3 FLOW DISTURBANCE AND LIQUID ALARMS Error message Alarm Warning LED Error FA-Warn: 1, 2 - x yellow Flow disturbance channels 1 and/or 2 LA-Alarm: 1, 2 x - red Liquid alarm channel 1and/or 2 LA-Break: 1, 2 x - red Cable break liquid sensor channel 1 and/or 2
fallen below => sensor defective
=> sensor defective
fallen below => sensor defective
For channels 2 and 3 appears when the control temperature is fallen below (+10° C alarm hysteresis)
of the following screen:
This prevents the heated line or the heated probe from overheating due to a defect ! The triggered
alarm (red LED) must be acknowledged by pressing any key. If the error is not corrected, the error
screen comes back after 60s.
In case of liquid alarm and activated LA-Reset (see 16.7) the following screen occurs :