Flow disturbance and liquid alarms – M&C TechGroup CSS-VC2 Operator's manual User Manual

Page 40

background image

40

Gas sampling and gas conditioning technology

14.1a-ME

17.2

TEMPERATURE ALARMS FROM COOLER AND TEMPERATURE CONTROL
MODULE

Error message

Alarm

Warning

LED

Error

Lowtemp.: 1

x

-

red

Low temperature cooler

Hightemp: 1

x

-

red

High temperature cooler

T> Max: 1

x

red

Temperature sensor range exceeded
=> sensor defective

T

x

red

Temperature sensor measuring range
fallen below => sensor defective

Lowtemp.: 2, 3

-

x

yellow Low temperature channel 2, 3

Hightemp.: 2, 3

x

-

red

High temperature channel 2, 3

T> Max.: 2, 3

x

red

Temperature sensor range exceeded
=> sensor defective

T

x

red

Temperature sensor measuring range
fallen below => sensor defective


For channels 2 and 3 appears when the control temperature is fallen below (+10° C alarm hysteresis)
of the following screen:


This prevents the heated line or the heated probe from overheating due to a defect ! The triggered
alarm (red LED) must be acknowledged by pressing any key. If the error is not corrected, the error
screen comes back after 60s.

17.3

FLOW DISTURBANCE AND LIQUID ALARMS

Error message

Alarm

Warning LED

Error

FA-Warn: 1, 2

-

x

yellow Flow disturbance channels 1 and/or 2

LA-Alarm: 1, 2

x

-

red

Liquid alarm channel 1and/or 2

LA-Break: 1, 2

x

-

red

Cable break liquid sensor channel 1 and/or 2


In case of liquid alarm and activated LA-Reset (see 16.7) the following screen occurs :