14.1a ME

Gas sampling and gas conditioning technology

35

Parameters:

Default setting

Info

Start in:

0d 0h 0m

Max.: 9T 23h 59m

Flushing time T1 30s

Max.: 99m 59s

Dead time

60s

0-120s

Interval T2

24h

On time T3

1s

Max.: 99s

Off time T4

5s

Max.: 99s

Pulse number

10

Is only calculated and cannot be <1

Activation

Off


16.4

DISPLAY OF EVENTS

By 4 x pressing (3 x without backflush module) of the arrow key

and then 1 x pressing of the PRG

key,

the event list appears in which the last 32 error messages and warnings are listed on 8

pages. The individual pages can be seen by using the arrow keys

.



At the bottom of the list of events, furthermore, the error-free time and the total run time (operating
time counter) are displayed. The error-free time indicates how much time has passed since the last
error. The total running time indicates how long the device has been running since the last service.

If service reset is performed, the error messages are erased and the times reset.

During a power failure, the data remains stored.