Electrical connections, Figure 5, Connecting heated line dn4/6 with antikink adapter – M&C TechGroup CSS-VC2 Operator's manual User Manual
Page 20
20
Gas sampling and gas conditioning technology
14.1a-ME
Mount the connection plate with 2 mounting screws on the rear panel
Mount the cover bracket with 2 connection screws on the device cover
12.3
CONNECTION OF THE HEATED LINE WITH ANTIKINK ADAPTER FOR
PORTABLE VERSION (ART. NO. 01G9060 OR 01G9061)
Sample gas IN
bulkhead fitting
Pressure ring 6mm Øi
Pressure ring 10mm Øi
Clip ring 10mm Øi
Clip ring 6mm Øi
Support sleeve 4mm
Øa
Junction nut 10mm Øi
Heated sample line
Adapter
Figure 5
Connecting heated line DN4/6 with antikink adapter
Place special adapters according to the above drawing on Teflon tubing;
Slide the support sleeve slide into the Teflon tubing;
Draw the Teflon tubing all the way into the bulkhead union 'Sample Gas ON' and hand-tighten
the adapter;
Tighten the adapter with wrench (SW 14) 1 1/4 turns; here hold the lock nut of the bulkhead
union with a wrench (SW 15);
Insert the 10mm pipe of the heating line all the way into the adapter and hand-tighten using the
nut;
Tighten nut with wrench (SW 19), 1 1/4 turns; here hold the adapter with wrench;
The union is now cut to be gas-tight and can be loosened as often as required.
12.4
ELECTRICAL CONNECTIONS
W A R N I N G !
Incorrect voltage can destroy the device. On connection, ensure
the correct mains voltage in accordance with the rating plate!
N O T I C E !
For the erection of power installations with nominal voltages up
to 1000V the requirements of VDE 0100 and relevant standards
and regulations must be complied with!