Electrical connections, Figure 5, Connecting heated line dn4/6 with antikink adapter – M&C TechGroup CSS-VC2 Operator's manual User Manual

Page 20

background image

20

Gas sampling and gas conditioning technology

14.1a-ME

Mount the connection plate with 2 mounting screws on the rear panel

Mount the cover bracket with 2 connection screws on the device cover



12.3

CONNECTION OF THE HEATED LINE WITH ANTIKINK ADAPTER FOR
PORTABLE VERSION (ART. NO. 01G9060 OR 01G9061)

Sample gas IN
bulkhead fitting

Pressure ring 6mm Øi

Pressure ring 10mm Øi

Clip ring 10mm Øi

Clip ring 6mm Øi

Support sleeve 4mm

Øa

Junction nut 10mm Øi

Heated sample line

Adapter

Figure 5

Connecting heated line DN4/6 with antikink adapter

Place special adapters according to the above drawing on Teflon tubing;

Slide the support sleeve slide into the Teflon tubing;

Draw the Teflon tubing all the way into the bulkhead union 'Sample Gas ON' and hand-tighten
the adapter;

Tighten the adapter with wrench (SW 14) 1 1/4 turns; here hold the lock nut of the bulkhead
union with a wrench (SW 15);

Insert the 10mm pipe of the heating line all the way into the adapter and hand-tighten using the
nut;

Tighten nut with wrench (SW 19), 1 1/4 turns; here hold the adapter with wrench;

The union is now cut to be gas-tight and can be loosened as often as required.



12.4

ELECTRICAL CONNECTIONS

W A R N I N G !


Incorrect voltage can destroy the device. On connection, ensure
the correct mains voltage in accordance with the rating plate!

N O T I C E !

For the erection of power installations with nominal voltages up
to 1000V the requirements of VDE 0100 and relevant standards
and regulations must be complied with!