Gas sampling and gas conditioning technology

14.1a-ME


After saving or discarding the settings, you will automatically return to the start screen.

If the unit is disconnected from the mains voltage while it is in user setup, all changes made are also
rejected.

The following values can be set:

Service interval

= (0 (-----) - 365 days/default setting 180Tage/off). After the set
time has elapsed, the warning LED (yellow) LIGHTS UP and
the following appears in plain text: "Service required". In
addition, the pending warning is also reported via the individual
alarm connection (see Section 12.4.2).

Language

= (German, English, French)

Cooler setpoint temperature = (0-7 ° C/default setting 5 ° C)

Temperature channel 2

= (temperature range dependent on sensor type (see below) /

default setting 180° C)

Temperature channel 3

= (temperature range dependent on sensor type (see below) /

default setting 180° C)

Thermocouple 2

= (T, N, K, J, E, ---/--- Default setting --- = PT100)

Thermocouple 3

= (T, N, K, J, E, ---/--- Default setting --- = PT100)

T : Cu/CuNi, -10 to +400°C
N : NiCrSi/NiSi, -50 to 800°C
K : NiCr/NiAl, -10 to +700°C
J : Fe/CuNi, 0 to +500°C
E : NiCr/CuNi, +10 to 400°C
--- : PT100, 0 to +240°C

LA-Reset = (x = reset necessary, ----- = no reset necessary)