Kipp&Zonen 2AP Sun Tracker User Manual
Page 86
REFERENCE
INFORMATION
85
ORigin
sensor. Returns OR. On all subsequent POsition commands, positions are given relative to this
user-defined origin. This command is only permitted after protection has been removed with the
PWord command.
PWord
Modification remains possible until 5 consecutive minutes elapse without communication. The
parameter
PWord
Disallow modification to configuration data. Returns PW.
PWord
match the four-digit number stored. The parameter
Modification must be enabled prior to entering this command.
SC
Retrieves the scaling factor for each sun sensor quadrant along with the nominal scaling of the
sensor. The scaling factor is used to convert the voltage from a given quadrant to the intensity that
is needed to create that voltage from that quadrant. The scaling of the sensor indicates the number
of degrees of error required to translate from "centred on the sun" to "no longer able to measure".
Returns SC
both azimuth and zenith.
SC
has been removed with the PWord command.
SErial
Retrieves the instrument serial number from permanent memory. Returns SE
SS
Retrieves the amount of time over which each sun observation takes place and the intensity level
below which the sun is deemed to be obscured. The sun must not be obscured at any time during
an observing period for an observation to be logged. Returns SP
SS
removed with the PWord command.
STeps
Retrieves from permanent memory the number of stepping units in a complete revolution for all
motors. Returns ST
STeps
number of stepping units depends on the setting for the motor controller. This command is only
permitted after protection has been removed with the PWord command.