TITRATION MODE

- allows you to save the graph as a bitmap file. Available only when the titration is

finished (after end point detection).

6.2

Titration Stop

The titration can be finished in one of the modes described below:

• Titration Completed. The titration was successfully terminated (with end point

successfully detected). This is the only mode with valid final result values.

• Manually Terminated. The current titration was manually terminated before end

point detection was achieved.

• Limits Exceeded. The preset maximum titrant volume was delivered without

reaching the end point. The titration is stopped with an error message.

• Critical Error. A critical error occurred and the titration was stopped. These

errors are normally related to the dosing system. The titration is stopped with a

specific error message.

• Potential Out of Range. The measured values from the input sensor are outside

the preset range (potential range). The titration is stopped with an error message.