Titration mode – Hanna Instruments HI 901 User Manual
Page 89
![background image](/manuals/737326/89/background.png)
6-5
TITRATION MODE
- allows you to save the graph as a bitmap file. Available only when the titration is
finished (after end point detection).
6.2
Titration Stop
The titration can be finished in one of the modes described below:
Titration Completed. The titration was successfully terminated (with end point
successfully detected). This is the only mode with valid final result values.
Manually Terminated. The current titration was manually terminated before end
point detection was achieved.
Limits Exceeded. The preset maximum titrant volume was delivered without
reaching the end point. The titration is stopped with an error message.
Critical Error. A critical error occurred and the titration was stopped. These
errors are normally related to the dosing system. The titration is stopped with a
specific error message.
Potential Out of Range. The measured values from the input sensor are outside
the preset range (potential range). The titration is stopped with an error message.