5-21

METHODS

The valid pre-titration volume is between the ranges shown below:

0.001 to 4.500 mL for a 5 mL burette

0.001 to 9.000 mL for a 10 mL burette

0.005 to 22.500 mL for a 25 mL burette

0.005 to 45.000 mL for a 50 mL burette

If no pre-titration volume is used, 0.000 mL should be entered.
Note: A pre-titration volume is highly recommended whenever possible. When fewer linear

doses are used the overall titration duration considerably shortens.

5.5.6

Pre-Titration Stir Time

When this option is enabled, the sample is mixed for a period of time before the titration

begins. This allows the sample to become homogeneous.

The range is from 0 to 180 seconds.

If 0 seconds is entered, the Pre-Titration Stir Time option is disabled.