Auxiliary functions – Hanna Instruments HI 901 User Manual
Page 106

8-6
AUXILIARY FUNCTIONS
8.3
Results
The Data Parameters screen contains options regarding data parameters from the titrator,
as follows:
8.3.1
Review Last Titration Report
Selecting this option, the last titration data can be reviewed.
The last titration report file can be displayed or printed. Also, the titration graph can be
reviewed by selecting
.
The information that can be reviewed in the report is based on the selections made in the
Setup Titration Report screen.
The following option keys are available:
Allows review of the last performed titration graph. The pH (or mV) versus titrant
volume curve can also be displayed. Also, if the Equivalence End Point option was
selected, the pH (or mV) curve and selected derivative curve (1
st
derivative, 2
nd
derivative) versus titrant volume are simultaneously displayed. Pressing
will change
the vertical axes scale units.
Allows printing the report to the printer.