beautypg.com

Methods – Hanna Instruments HI 901 User Manual

Page 75

background image

5-35

METHODS

If the titrant 1 unit is selected as M (mol/L), titrant 2 unit is selected as M (mol/L) and the

final result unit as mg/g (milligrams/gram) the following formula is used to calculate the

amount of titrant 1 (used in the first stage of back titration or direct titration) and will be as

follows:

The formula is based on an approximation: the sample concentration is 100% w/w.

With this approximation, the titrator will calculate the volume of titrant 1 needed to consume

the sample and multiply it with an excess factor in order to raise or to lower the amount of

titrant 1 dispensed.

Some variables can be set according to the amount of sample and titrant 1 used.
When the parameters from the formula are set, press

to proceed with the next

formula used in the back titration algorithm.

If you do not want to use the “Calc. Direct Titr. Volume” formula, then from the View / Modify

Method screen:

• highlight the Titrant 1 Entry option, press

.

• highlight Fixed Weight or Volume option and press

, you will be prompted to

type the volume of titrant 1 to be dispensed in the first stage of back titration.

This formula is used to calculate the remaining volume of titrant 1 after the reaction with the

sample.