Methods – Hanna Instruments HI 901 User Manual
Page 53

5-13
METHODS
5.5.4End Point Mode
5.5.4.1 Fixed End Point (pH or mv)
Fixed End Point (pH):
The titration is normally terminated when the preset pH value has been exceeded. The
reported end point volume is interpolated between the dispensed volume when pH is under
the preset pH value and the dispensed volume when pH exceeded the preset pH value.
The range is from - 2.000 to 20.000 pH.