Methods – Hanna Instruments HI 901 User Manual

Page 63

background image

5-23

METHODS

The signal stability window (condition) represents the time interval

t during which the

potential measured in solution (mV) is confined inside the potential interval

E (see the

boxes on the chart).

The new signal value is acquired if the stability condition is reached but not before the

minimum waiting time (t min wait) has elapsed.

If the t max wait time has elapsed and the stability condition is not reached, the potential is

acquired (and a new dose is added).

delta E - is the height of the signal stability window.

The range is from 0.1 to 99.9 mV.

delta t -is the width of the signal stability window.

The range is from 0.5 to 10.0 seconds.

t min wait - is the minimum elapsed time before stability check. This is also the minimum

elapsed time between two doses.
The range is from 2 seconds to t max wait time.

t max wait - is the maximum elapsed time between two successive doses. If the t max wait

has elapsed, a new dose is added even if the condition of signal stability is not

reached.

The range is from t min wait time to 180 seconds.