To attach the burette to the pump simply slide the burette into the pump support as shown

in the figure presented in 9.1.2 Burette Changing section.

9.1.5

Cleaning Burette

To clean the burette, follow these steps:

• If the burette is filled with liquid, remove the aspiration tube from the titrant bottle

and execute purge burette operation (see 8.1.4 Purge Burette section).

• Insert the aspiration tube into cleaning solution and/or into deionized water or

titrant solvent, depending on the chemical nature of titrant and the intended use of

the burette. If a titration will follow, use titrant solvent to clean the tube.

• Execute prime burette operation to fill the burette (use 2 rinses) (see 8.1.1 Prime

Burette section).

• During second refilling of the burette remove the aspiration tube out of the solvent

beaker and allow the air to replace the liquid in the burette. This will clean the

aspiration tube.

If this simple cleaning procedure is not adequate for the desired purpose, continue with

these steps:

• Slide the burette out from the pump assembly.
• Remove the tubes with the special mounting tool (tool for valve fitting and burette

cap removal).

• Clean them separately or insert new ones.
• Remove the protective cap from the bottom by using the special tool.
• Remove the protective plastic shield.
• Remove the syringe from the burette assembly by unscrewing it with your fingers.
• Extract the piston from the syringe.