5-19

METHODS

Range:

Range is an optional feature for equivalence point recognition. It represents a set mV or pH

range for equivalence point detection.

The Range option can be enabled by selecting YES in the Range Options screen.

Limit 1 - The first limit of the range attained during the titration:

This range can be from -2.000 pH to 20.000 pH (for pH equivalence point).

This range can be from -2000.0 mV to 2000.0 mV (for mV equivalence point).

Limit 2 - The second limit of the range attained during titration:

This range can be from -2.000 pH to 20.000 pH (for pH equivalence point).

This range can be from -2000.0 mV to 2000.0 mV (for mV equivalence point).

The Limit 2 value must not be equal to the Limit 1 value.