5.5.22 Signal Averaging

This option enables or disables a filtering on the mV/pH reading.

If 1 Reading is selected, the filtering is disabled. Otherwise, the titrator will take the last reading

and place it into a “moving window” along with the last 2, 3 or 4 readings (depending on the

selected option). The average of those readings is displayed and used for calculations.

Averaging more readings is helpful when a noisy signal is received from the electrode.

5.6 Printing

To print method parameters, press

from the main screen.

Press

and wait a few seconds until the printer completes the job.

If no printer is connected to the dedicated socket, or if the printer is offline, an error message

will appear on the display (see 9.3.3 Connecting a Printer section, for information about
connecting a printer to the titrator).