beautypg.com

Appendix 3, A3 multiple equivalence points (hi 902 only) – Hanna Instruments HI 901 User Manual

Page 135

background image

A3-3

APPENDIX 3

A3

MULTIPLE EQUIVALENCE POINTS (HI 902 only)

A3.1

Applicability Domain

In case of:

• determining the components concentration from a synthetic mixture (e.g. Mixture of

HCl, CH

3

COOH, NH

4

Cl).

• determining the concentration of a polyprotic acid in it’s titrable ionization stages

(e.g. H

3

PO

4

, two equivalence points in aquogenos medium).

A3.2

Perform a Multi-Equivalence Point Titration

With the information provided above, set up the method parameters as follows:

• Set the end point mode as equivalence end point (mV or pH).
• Set the number of equivalence points (up to 5 points) to be found.

• Select the end point determination (like in standard titration).

• Set the rest of method parameters and save the modification.

This manual is related to the following products: