MAINTENANCE, PERIPHERALS

9.2

Fuses Replacement

To replace the fuses, follow these steps:

• Turn off the titrator.
• Remove the power cord from the power connector located on the rear side of the

titrator case.

• With a screw driver open the fuse holder lid.

• Extract the fuse holder.
• Replace the fuses (for recommended fuses replacement please see the label located

above the main power switch).

• Close the fuse holder lid.
• Connect the power cord.

Note: For other maintenance operations please contact your dealer or the nearest Hanna

Customer Service Center.

9.3 Peripherals

Warning! Connection/disconnection of POWER CORD, PUMP ASSEMBLY, EXTERNAL PC

DISPLAY, PRINTER, RS232 INTERFACE, EXPANSION DEVICE must be done

only when titrator and external devices are turned off.