5-36

METHODS

In order to calculate this volume, the remaining titrant 1 volume is titrated (pH/mV equivalence

point titration) with titrant 2.

After the excess volume of titrant 1 is calculated, the following formula is used to calculate

the exact volume of titrant 1 that was consumed by the sample:

V1 = V1tot – V1excess

When all the parameters are set, press

to proceed with the “Calculating Sample

Concentration” formula:

Sample Calculations by Volume
When choosing formula, follow these steps:

First select the titrant 1 unit