TITRATION MODE
6
TITRATION MODE
6.1
Titration Start
Before beginning to perform a titration make sure that the following conditions are met:
A pump is properly installed and selected as active.
A burette is inserted in the pump and filled with titrant.
The aspiration tube is inserted in the titrant bottle and the dispensing tube is over
the analyte beaker.
The electrode and the temperature probe is inserted in the analyte beaker.
The desired method is selected as active and the parameters are set at optimum
values.
6.1.1
In Progress Titration
To start a new titration, press
from the main screen.
When a titration begins:
The stirrer will turn on (if detected and enabled).
If the pre-stirring time option is enabled, the sample will be stirred until the
prescribed time elapses (see 5.5.6 Pre-Titration Stir Time).
If the pre-titration volume option is enabled, the prescribed volume will be dis-
pensed (see 5.5.5 Pre-Titration Volume).
According to the Measurement Mode and the Dosing Type option, the titrator will
start to deliver doses until the titration end point are detected or a titration stop
condition occurs.
6.1.2
Suspend Titration
While titration is in progress, you can temporarily stop it by pressing
. All the titration
parameters will be frozen.
You can continue the titration by pressing
.
6.1.3
On-line Graph
During a titration, both the potentiometric S-shaped curve and the selected derivative curve
(titration with equivalence point only) can be displayed on the Titration Graph screen, by
pressing
. The titration ID report is also displayed inside the graph window.