6-3

TITRATION MODE

6

TITRATION MODE

6.1

Titration Start

Before beginning to perform a titration make sure that the following conditions are met:

• A pump is properly installed and selected as active.
• A burette is inserted in the pump and filled with titrant.
• The aspiration tube is inserted in the titrant bottle and the dispensing tube is over

the analyte beaker.

• The electrode and the temperature probe is inserted in the analyte beaker.
• The desired method is selected as active and the parameters are set at optimum

values.

6.1.1

In Progress Titration

To start a new titration, press

from the main screen.

When a titration begins:

• The stirrer will turn on (if detected and enabled).
• If the pre-stirring time option is enabled, the sample will be stirred until the

prescribed time elapses (see 5.5.6 Pre-Titration Stir Time).

• If the pre-titration volume option is enabled, the prescribed volume will be dis-

pensed (see 5.5.5 Pre-Titration Volume).

• According to the Measurement Mode and the Dosing Type option, the titrator will

start to deliver doses until the titration end point are detected or a titration stop

condition occurs.

6.1.2

Suspend Titration

While titration is in progress, you can temporarily stop it by pressing

. All the titration

parameters will be frozen.

You can continue the titration by pressing

.

6.1.3

On-line Graph

During a titration, both the potentiometric S-shaped curve and the selected derivative curve

(titration with equivalence point only) can be displayed on the Titration Graph screen, by
pressing

. The titration ID report is also displayed inside the graph window.