5-30

METHODS

M (mol/L)

(Molarity; moles/liter)

N (eq/L)

(Normality; equivalences/liter)

mg/L

(milligrams/liter)

µg/L

(micrograms/liter)

mmol/L

(millimoles/liter)

mg/mL

(milligrams/milliliter)

g/100 mL

(grams/100 milliliters)

eq/L

(equivalences/liter)

meq/L

(milliequivalences/liter)

If the titrant unit is selected as N (eq/L) and the final sample unit as g/L (grams/liter), the

formula used to generate the result is displayed below.

Some variables can be set according to the amount of sample and titrant used.

Standardize Titrant by Weight
This calculation should be used when the titrant is standardized (concentration

determination) using a solid standard. Determination of the titrant concentration is based

on the primary standard weight (in grams).