background image

5-32

METHODS

eq/L

(equivalences/liter)

meq/L

(milliequivalences/liter)

The formula can be used either for titrant standardization or for sample analysis.

C*V*F1*F2*F3

S

Where:

C = the concentration of the titrant
F1 = general factor
F2 = general factor
F3 = general factor
S = sample size, in grams or milliliters
V = the volume delivered, in liters, to reach the preset or equivalence end point

(determined by the titrator)