beautypg.com

Hanna Instruments HI 99131 User Manual

Instruction manual, Portable waterproof ph meter for plating baths

background image

Instruction Manual

HI 99131

Portable Waterproof

pH Meter for Plating Baths

Dear Customer,
Thank you for choosing a Hanna Instruments
product.
Please read this instruction manual care-
fully before using the instrument. If you need
additional technical information, do not hesi-
tate to e-mail us at [email protected]
This instrument is in compliance with the

directives.

All Hanna Instruments meters are warranted
for two years
against defects in workmanship
and materials when used for their intended pur-
pose and maintained according to instructions.
The probes are warranted for a period of
six months.

This warranty is limited to repair or replacement
free of charge. Damages due to accident, mis-
use, tampering or lack of prescribed maintenance
are not covered.

If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of pur-
chase, serial number and the nature of the prob-
lem. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instru-
ments, first obtain a Returned Goods Authoriza-
tion Number from the Customer Service depart-
ment and then send it with shipment costs pre-
paid. When shipping any instrument, make sure it
is properly packaged for complete protection.

All rights are reserved. Reproduction in whole or
in part is prohibited without the written consent of
the copyright owner.

WARRANTY

PRELIMINARY EXAMINATION

Remove the instrument from the packing
material and examine it carefully to make
sure that no damage has occurred during
shipment. If noticeable damage is evident,
notify your dealer.

Note: Save all packing material until you

are sure that the instrument func-
tions correctly. All defective items
must be returned in the original
packing together with the supplied
accessories.

HI 99131 is a portable, microprocessor-
based pH and temperature meter with a
special probe.
The HI 62911 pH electrode has been spe-
cially designed and fully shielded for appli-
cations in plating baths. The titanium elec-
trode body acts like a Faraday cage, and
allows stable readings even in samples
where strong electrical fields are involved.
Moreover, a built-in temperature sensor al-
lows simultaneous pH and temperature
readings and a mini amplifier renders mea-
surements impervious to noise and elec-
trical interferences.

Each meter is supplied complete with:

HI 62911 pH/temperature probe

calibration & cleaning solution sachets

3x1.5V AA alkaline batteries

instruction manual

rugged carrying case

GENERAL DESCRIPTION

SPECIFICATIONS

Range

-2.00 to 16.00 pH

-5.0 to 105.0°C / 23.0 to 221.0°F

Resolution

0.01 pH

0.1°C / 0.1°F

Accuracy

(@20°C)

±0.02 pH

±0.5°C up to 60°C; ±1°C outside

±1°F up to 140°F; ±2°F outside

Typical EMC

±0.02 pH

Deviation

±0.2°C or ±0.4°F

Temperature

Automatic

Compensation

pH Calibration

Automatic, 1 or 2 point

with 2 sets of memorized buffers

(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)

Probe

(included)

HI 62911

amplified pH probe with built-in

temperature sensor and titanium cage

working as matching pin

Battery Type

3 x 1.5V AA / IEC LR6

Battery Life

Approx. 1500 hours

Auto-off

After 8 minutes of non-use

Environment

0 to 50°C (32 to 122°F);

RH max 100%

Dimensions

150x80x36 mm (5.9x3.1x1.4”)

Weight

210 g (7.4 oz.)

To clean the meter, it is recommended to use water.

HI 62911

Amplified pH electrode with
built-in temperature sensor,
DIN connector & titanium cage
working as matching pin

HI 7004L

pH 4.01 buffer solution, 500 mL

HI 7006L

pH 6.86 buffer solution, 500 mL

HI 7007L

pH 7.01 buffer solution, 500 mL

HI 7009L

pH 9.18 buffer solution, 500 mL

HI 7010L

pH 10.01 solution, 500 mL

HI 70300L

Storage solution, 500 mL

HI 7061L

Cleaning solution, 500 mL

HI 710007

Blue shockproof rubber boot

HI 710008

Orange shockproof rubber boot

HI 721312

Rugged carrying case

HI 76405

Electrode holder

ACCESSORIES

LCD DESCRIPTION

1. Automatic Temperature Compensation
indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display

IST99131R3 07/05

w w w . h a n n a i n s t . c o m

Hanna Instruments reserves the right to

modify the design, construction and appear-

ance of its products without advance notice.