Methods – Hanna Instruments HI 901 User Manual
Page 54

5-14
METHODS
Fixed End Point (mV):
The end point detection algorithm is the same as for pH, but the threshold value is expressed
in mV.
The range is from - 2000.0 to 2000.0 mV.
5.5.4.2
Equivalence End Point (pH or mv)
The titration is normally terminated when the equivalence point is detected (the point where
the added quantity of titrant equals the quantity of analyte present in the sample, according
to the chemical combination ratio).