SCPI Command Reference

92

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:CURRent:ACCBase

Description

This command is used to set the ACC base

current. The base current is the actual current of

the laser. For this command to function properly,

the instrument must be in High Wavelength

Stability or On/Off modulation mode and the

SOURce[1..n]:STABle? query must return 1

(stable). Normally, this command would rarely

be used.

*RST has no effect on this command.

Syntax

:SOURce[1..n]:CURRent:ACCBase

Parameter(s)

ACCBase:
The program data syntax for is

defined as a

DATA> element.

The parameter is the desired

current of the instrument in amperes.

Example(s)

SOUR:PULM:STATE OFF

SOUR:WAV:MODE HRES

/* Under FDA jurisdiction, a password must be

entered before turning the laser on. */

SOUR:POW:PROT:SPWD "safekey"

SOUR:POW:STAT ON