beautypg.com

Source[1..n]:pulm:internal:depth – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 126

background image

SCPI Command Reference

120

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:PULM:INTernal:DEPTh

Description

This command is used to set the dither

modulation signal depth. If the instrument is not

in Dither modulation mode, this command will

have no effect.

At *RST, the dither depth is 0.005 amperes.

Syntax

:SOURce[1..n]:PULM:INTernal:DEPTh

Parameter(s)

DepthModulation:
The program data syntax for

is defined as a

NUMERIC PROGRAM DATA> element.

The parameter represents

the new dither modulation signal depth in

amperes.

Example(s)

SOUR:PULM:STAT ON

SOUR:WAV:MODE NORMAL

SOUR:PULM:INT:DEPT 0.00255

See Also

SOURce[1..n]:PULM:INTernal:DEPTh?

SOURce[1..n]:PULM:LIMit:DEPTh:HIGH?

SOURce[1..n]:PULM:LIMit:DEPTh:LOW?

SOURce[1..n]:PULM:LIMit:DEPTh:STEP?