beautypg.com

Source[1..n]:accoutput – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 92

background image

SCPI Command Reference

86

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:ACCOutput

Description

This command sets the delta temperature and

the delta current of the laser.

At *RST, the delta current and delta temperature

are device-dependent.

Syntax

:SOURce[1..n]:ACCOutput

,

Parameter(s)

³

DeltaTemp:

The program data syntax for is

defined as a

DATA> element.

The parameter represents the

delta temperature of the laser in °C.

The value used must be equal to or between the

values returned by

SOURce[1..n]:TEMPerature:PROTection:LOW

[:LEVel]? and

SOURce[1..n]:TEMPerature:PROTection:HIGH

[:LEVel]?. The query

SOURce[1..n]:TEMPerature:PROTection:STEP?

returns the instrument sensitivity.