:SOURce[1..n]:PULM:INTernal:DEPTh?

Description

This query returns the dither modulation signal

depth. If the instrument is not in Dither

Modulation mode, the returned value is

undefined.

At *RST, the dither depth is 0.005 amperes.

Syntax

:SOURce[1..n]:PULM:INTernal:DEPTh?

Parameter(s)

None

Response Syntax

Response(s)

DepthModulation:
The response data syntax for

is defined as a

NUMERIC RESPONSE DATA> element.

The response is a value

representing the current dither modulation signal

depth in amperes.

Example(s)

SOUR:PULM:STAT ON

SOUR:WAV:MODE NORMAL

SOUR:PULM:INT:DEPT 3.15E-3

SOUR:PULM:INT:DEPT? Returns 3.150000E-003

See Also

SOUR:PULM:INT:DEPT

SOUR:PULM:LIM:DEPT:HIGH

SOUR:PULM:LIM:DEPT:LOW?

SOUR:PULM:LIM:DEPT:STEP?