beautypg.com

Source[1..n]:pulm:limit:depth: delta – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 138

background image

SCPI Command Reference

132

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:PULM:LIMit:DEPTh:

DELTa?

Description

This query returns the minimum step available

when changing the dither modulation signal

depth with the

SOURce[1..n]:PULM:INTernal:DEPTh command.

*RST has no effect on this command.

Syntax

:SOURce[1..n]:PULM:LIMit:DEPTh:DELTa?

Parameter(s)

None

Response Syntax

Response(s)

DepthPrecision:
The response data syntax for

is defined as a

DATA> element.

The response is the

minimum step available for the dither

modulation signal depth in amperes.

Example(s)

SOUR:PULM:LIM:DEPT:DELT? Returns

1.000000E-004

Notes

The step is an indication of the instrument

sensitivity. An attempt to change the dither

modulation signal depth by a quantity less than

the minimum step will be ignored.

See Also

SOURce[1..n]:PULM:INTernal:DEPTh

SOURce[1..n]:PULM:INTernal:DEPTh?

SOURce[1..n]:PULM:LIMit:DEPTh:HIGH?

SOURce[1..n]:PULM:LIMit:DEPTh:LOW?

SOURce[1..n]:PULM:LIMit:DEPTh:STEP?