SCPI Command Reference

WDM Laser Source

83

Product-Specific Commands—Description

Product-Specific Commands—Description

:CALibration[1..n]:VALue

Description

This command is used to set the power

calibration value of the instrument. This

calibration value is used as a correction factor.

At *RST, the calibrated power is set to the factory

value.

Syntax

:CALibration[1..n]:VALue

DBM|W]>

Parameter(s)

CalibrationValue:
The program data syntax for

is defined as a

DATA> element followed by an optional

element. The

allowed elements

are: DBM|W.

The parameter represents

the new calibration value. If you do not specify a

unit, the device will use the current power units

(Watts or dBm). To select the current power units

use the UNIT[1..n]:POWer command.

Example(s)

SOUR:PULM:STAT OFF

SOUR:WAV:MODE NORMAL

/* Under FDA jurisdiction, a password must be

entered before turning the laser on. */

SOUR:POW:PROT:SPWD "safekey"

SOUR:POW:STAT ON